2 deaths reported in recall of 6 million Teleflex endotracheal tubes

June 21, 2019 By Nancy Crotti

Teleflex is recalling this 6 mm endotracheal tube in a Class 1 recall. (Image from Teleflex)

Teleflex (NYSE:TFX) is recalling multiple models of its Hudson RCI Sheridan and Sheridan endotracheal tubes due to complaints of the Sheridan connector disconnecting from the breathing circuit. Two deaths and one injury have been reported to the company, according to the FDA.

Each of the more than 6 million recalled tracheal tubes includes an appropriately sized 15mm connector. Although the connector can be removed, it should not disconnect from the endotracheal tube without significant force, the agency said. The FDA has termed it a Class 1 recall, the most serious type of recall involving devices that may cause injury or death.

Teleflex (Wayne, Penn.) has received 179 complaints with 192 occurrences in association with the endotracheal tube disconnection, according to the FDA. In each case of patient death, a decision was made to extubate and reintubate the patient, according to the field notice that Teleflex issued to providers May 27, 2019. The company has began selling the affected tubes in October 2016.

“While initially successful, the patients expired in the presence of subsequent reintubation attempts and other unrelated co-morbidities,” the company’s letter said. “In most reported cases, detachment of the connector was identified by clinical personnel or via eventual decrease in ventilator circuit pressure which triggered ventilator alarms.”

The field notice advises clinical staff to ensure that the 15 mm connector is seated firmly in any of the affected endotracheal tubes already in place to prevent disconnection. Should disconnection occur, the company is advising reconnecting the two components promptly and securely in the manner described in the instructions for use or consider replacing the connector, making sure to evaluate the risks associated with reintubation.

Certain Sheridan-branded EIF-000361 endotracheal tubes and connectors in sizes 6.0, 6.5, 7.0, 7.5, 8.0 and 8.5 mm are included in this recall. Specific lots are:

    • Hudson RCI Sheridan LTS.
    • Hudson RCI Sheridan (CF, Preformed, HZT) 6.0, 6.5, 7.0, 7.5, 8.0, 8.5 mm.
    • Hudson RCI Sheridan/EZ-ENDO 6.0, 6.5, 7.0, 7.5, 8.0, 8.5 mm.
    • Hudson RCI Sheridan Uncuffed 6.0, 6.5, 7.0 mm.
    • Sheridan/CF Novaplus 7.0.
    • Sheridan/HVT Novaplus 7.0, 7.5, 8.0, 8.5 mm.

A full list of the recalled devices is available here.

Teleflex stock dipped nearly 1% to $330.22 on news of the recall this morning.

Filed Under: Tubing Components Tagged With: fda, teleflexincorporated

Primary Sidebar

DeviceTalks Tuesdays

DeviceTalks Tuesdays
MDO ad

Stay Current

Need Medical Tubing news in a minute?
We Deliver!

Medical Tubing + Extrusion newsletter get you caught up on all the mission critical news you need in medical tubing. Sign up today.

Enews Sign up

Sponsored Content

A new way to access scientific papers?

Mass Device

The Medical Device Business Journal. MassDevice is the leading medical device news business journal telling the stories of the devices that save lives.

Visit Website

Footer

Inv Logo

MASSDEVICE MEDICAL NETWORK

MassDevice
DeviceTalks
Medical Tubing & Extrusion
Medical Design & Outsourcing
MedTech 100 Index
Drug Discovery & Development
Pharmaceutical Processing World
Medical Design Souring
R&D World

Medical Tubing & Extrusion

Subscribe to our e-newsletter
Advertise with us
About
Attend our Monthly Webinars
Listen to our Weekly Podcasts
Join our Device Talks Tuesdays Discussion
Add us on FacebookFollow us on TwitterConnect with us on LinkedIn Follow us on YouTube

Copyright © 2022 · WTWH Media LLC and its licensors. All rights reserved.
The material on this site may not be reproduced, distributed, transmitted, cached or otherwise used, except with the prior written permission of WTWH Media.

Advertise | Privacy Policy | RSS